ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

原資料は米国FDAの発表(英語)です。商品名・理由は原文のまま表記します。

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

製造・販売社Erbe USA Inc
リコール理由A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
リスク等級Class II (Class II · 中)
進行状況Ongoing
報告日2026年8月19日
リコール開始日2026年6月26日
地域GA, United States
流通範囲US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.
数量11226 units
コード/ロットUDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
区分Voluntary: Firm initiated
リコール番号Z-2938-2026

公式出典

U.S. FDA — Recalls, Market Withdrawals & Safety Alerts fda.gov ↗

リコール番号でFDA公式サイトから再確認できます。 (リコール番号: Z-2938-2026)