← Back to recall lookup
Device Class II · moderate Ongoing

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

Recalling firmErbe USA Inc
ReasonA strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Risk classClass II (Class II · moderate)
StatusOngoing
Report dateAug 19, 2026
InitiatedJun 26, 2026
RegionGA, United States
DistributionUS Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.
Quantity11226 units
Code/LotUDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
TypeVoluntary: Firm initiated
Recall numberZ-2938-2026

Official source

U.S. FDA — Recalls, Market Withdrawals & Safety Alerts fda.gov ↗

You can re-verify each item on the official FDA site using its recall number. (Recall number: Z-2938-2026)

openFDA (U.S. Food & Drug Administration) — food/drug/device enforcement · CC0 1.0 (public domain)

Find out more

The links above open external searches by recall number / product name; the results are the responsibility of each site.

This page summarizes U.S. FDA data for general information and is not medical advice. Confirm details via official FDA announcements and the manufacturer.