Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
원자료는 미국 FDA 발표(영문)입니다. 상품명·사유는 원문 그대로 표기합니다.
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
| 제조·판매사 | Medtronic MiniMed, Inc. |
|---|---|
| 리콜 사유 | Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery. |
| 위해 등급 | Class II (Class II · 중간) |
| 진행 상태 | Ongoing |
| 보고일 | 2026년 10월 14일 |
| 리콜 개시일 | 2026년 8월 20일 |
| 지역 | CA, United States |
| 유통 범위 | U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV. |
| 수량 | 25 |
| 코드/로트 | MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB000157… |
| 구분 | Voluntary: Firm initiated |
| 리콜 번호 | Z-0088-2027 |
공식 출처
U.S. FDA — Recalls, Market Withdrawals & Safety Alerts fda.gov ↗리콜 번호로 FDA 공식 사이트에서 다시 확인할 수 있습니다. (리콜 번호: Z-0088-2027)
openFDA (U.S. Food & Drug Administration) — food/drug/device enforcement · CC0 1.0 (public domain)
관련 정보 더 찾기
위 링크는 리콜 번호·제품명으로 외부 검색을 여는 것이며, 검색 결과의 내용은 각 매체·사이트의 책임입니다.
이 페이지는 미국 FDA 데이터를 정리한 생활정보이며 의료 조언이 아닙니다. 정확한 내용은 FDA 공식 발표와 제조사 안내로 확인하세요.