VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006
원자료는 미국 FDA 발표(영문)입니다. 상품명·사유는 원문 그대로 표기합니다.
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
| 제조·판매사 | Steris Corporation |
|---|---|
| 리콜 사유 | Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction |
| 위해 등급 | Class II (Class II · 중간) |
| 진행 상태 | Ongoing |
| 보고일 | 2026년 10월 14일 |
| 리콜 개시일 | 2026년 8월 31일 |
| 지역 | OH, United States |
| 유통 범위 | Worldwide - US Nationwide distribution. |
| 수량 | 266 |
| 코드/로트 | UDI: 10724995220324 Lot numbers: 260810 - expiration date 08/10/2026 260810A - expiration date 08/10/2026 261106 - expiration date 11/06/2026 270116 - expiration date 01/16/2027 |
| 구분 | Voluntary: Firm initiated |
| 리콜 번호 | Z-0070-2027 |