← リコール検索へ
医療機器 Class II · 中 Ongoing

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

原資料は米国FDAの発表(英語)です。商品名・理由は原文のまま表記します。

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

製造・販売社Steris Corporation
リコール理由Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
リスク等級Class II (Class II · 中)
進行状況Ongoing
報告日2026年10月14日
リコール開始日2026年8月31日
地域OH, United States
流通範囲Worldwide - US Nationwide distribution.
数量2205
コード/ロットUDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiratio…
区分Voluntary: Firm initiated
リコール番号Z-0073-2027

公式出典

U.S. FDA — Recalls, Market Withdrawals & Safety Alerts fda.gov ↗

リコール番号でFDA公式サイトから再確認できます。 (リコール番号: Z-0073-2027)

openFDA (U.S. Food & Drug Administration) — food/drug/device enforcement · CC0 1.0 (public domain)

関連情報をさらに探す

上のリンクはリコール番号・商品名で外部検索を開くものであり、検索結果の内容は各サイトの責任です。

このページは米国FDAのデータを整理した生活情報であり、医療アドバイスではありません。正確な内容はFDA公式発表と製造社の案内でご確認ください。