← Back to recall lookup
Device Class II · moderate Ongoing

Medline Micro-Kill Bleach Germicidal Bleach Wipes,7" x 8", 150 count, 150 count, Product Number MSC351400AN

Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.

Recalling firmMedline Industries, LP
ReasonMedline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.
Risk classClass II (Class II · moderate)
StatusOngoing
Report dateAug 19, 2026
InitiatedJun 30, 2026
RegionIL, United States
DistributionUS Nationwide distribution in the states and territories of Guam, Puerto Rico.
Quantity591,561 units
Code/LotUDI-DI (each) 10888277652514, UDI-DI (case) 20888277652511, Lot Numbers: All lots produced between 05/20/2025 and 05/8/2026
TypeVoluntary: Firm initiated
Recall numberZ-2923-2026

Official source

U.S. FDA — Recalls, Market Withdrawals & Safety Alerts fda.gov ↗

You can re-verify each item on the official FDA site using its recall number. (Recall number: Z-2923-2026)

openFDA (U.S. Food & Drug Administration) — food/drug/device enforcement · CC0 1.0 (public domain)

Find out more

The links above open external searches by recall number / product name; the results are the responsibility of each site.

This page summarizes U.S. FDA data for general information and is not medical advice. Confirm details via official FDA announcements and the manufacturer.