← Back to recall lookup
Device Class II · moderate Ongoing

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Recalling firmSteris Corporation
ReasonGeobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Risk classClass II (Class II · moderate)
StatusOngoing
Report dateOct 14, 2026
InitiatedAug 31, 2026
RegionOH, United States
DistributionWorldwide - US Nationwide distribution.
Quantity1185
Code/LotUDI: 10724995023833 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiratio…
TypeVoluntary: Firm initiated
Recall numberZ-0072-2027

Official source

U.S. FDA — Recalls, Market Withdrawals & Safety Alerts fda.gov ↗

You can re-verify each item on the official FDA site using its recall number. (Recall number: Z-0072-2027)

openFDA (U.S. Food & Drug Administration) — food/drug/device enforcement · CC0 1.0 (public domain)

Find out more

The links above open external searches by recall number / product name; the results are the responsibility of each site.

This page summarizes U.S. FDA data for general information and is not medical advice. Confirm details via official FDA announcements and the manufacturer.